Initial assessment of tumor tracking with a gimbaled linac system in clinical circumstances: a patient simulation study.

Depuydt T, Poels K, Verellen D, Engels B, Collen C, Haverbeke C (+10 more)

Abstract

PurposeTo have an initial assessment of the Vero Dynamic Tracking workflow in clinical circumstances and quantify the performance of the tracking system, a simulation study was set up on 5 lung and liver patients.Methods and materialsThe preparatory steps of a tumor tracking treatment, based on fiducial markers implanted in the tumor, were executed allowing pursuit of the tumor with the gimbaled linac and monitoring X-rays acquisition, however, without activating the 6 MV beam. Data were acquired on workflow time-efficiency, tracking accuracy and imaging exposure.ResultsThe average time between the patient entering the treatment room and the first treatment field was about 9 min. The time for building the correlation model was 3.2 min. Tracking errors of 0.55 and 0.95 mm (1σ) were observed in PAN/TILT direction and a 2D range of 3.08 mm. A skin dose was determined of 0.08 mGy/image, with a source-to-skin distance of 900 mm and kV exposure of 1 mAs. On average 1.8 mGy/min kV skin dose was observed for 1 Hz monitoring.ConclusionThe Vero tracking solution proved to be fully functional and showed performance comparable with other real-time tracking systems.