Radiation therapy with unflattened photon beams: dosimetric accuracy of advanced dose calculation algorithms.

Kragl G, Albrich D, Georg D.

Abstract

Purpose To compare the dosimetric accuracy of advanced dose calculation algorithms for flattened (FF) and unflattened (FFF) photon beams. Material and methods We compared the enhanced collapsed cone (eCC) algorithm implemented in OncentraMasterplan and the XVMC (MC) code in Monaco. Test plans were created for 10MV FF and FFF beams. Single beam tests were delivered to radiochromic films positioned within a solid water phantom and evaluated with 1D γ-index analysis. Conformal plans were verified with ion chambers in an anthropomorphic thorax phantom. IMRT plans were applied to the Delta4 system and evaluated with γ-criteria of 3% and 3mm. Results 1D γ-index evaluation revealed significantly lower (p Conclusions The dose calculation accuracy for unflattened beams was found to be at least as high as for flattened beams. The slightly improved dose calculation accuracy observed for off-axis profiles for single FFF beams did not directly translate into better verification results for composite IMRT plans.