Patient Cranial Angle and Intrafractional Stability in CyberKnife Robotic Radiosurgery: A Retrospective Analysis.

Kang CL, Huang YY, Chen YR, Tsai SH, Huang CC, Huang YJ.

Abstract

Purpose: The aim of this study was to investigate whether variations in cranial angles and treatment accuracy during CyberKnife robotic radiosurgery necessitate adjustment of the margins of the planning target volume. Patients and Methods: Data from 66 patients receiving CyberKnife treatment for brain tumors were retrospectively analyzed. Patients were immobilized using a thermoplastic mask and headrest. The cranial angle was measured on planning CT and patients were divided into 2 groups: ≤10° (Group A) and >10° (Group B). Intrafractional motion was recorded using the CyberKnife tracking system over 50 min. Translational and rotational errors were compared between groups, and planning target volume margins were calculated. Results: In Group A, significant translational error differences were found along with the X-axis over time (P P P P Conclusion: Adjusting cranial angle to ≤10° during thermoplastic mask molding provided better or similar intrafractional stability compared to >10°.