Impact of field number and monitor units per segment on magnetic resonance-guided hypofractionated stereotactic radiotherapy for brain metastases: plan quality, deliverability, and robustness trade-offs.

Ren S, He J, Han S, Xu N, Wang W, Zhu S (+5 more)

Abstract

BackgroundMagnetic resonance-guided linear accelerators (MR-Linacs) have recently been introduced for radiotherapy of brain metastases (BMs), including hypofractionated stereotactic radiotherapy (HSRT). However, optimal strategies for planning HSRT within MR-guided adaptive workflows remain to be established. This study aimed to evaluate the influence of field number and minimum monitor unit per segment (MU/segment) on plan quality, and robustness of dose calculations in MR-guided HSRT for patients with solitary medium-sized BMs.MethodsThis retrospective study included 20 patients who underwent HSRT for solitary medium-sized BMs, receiving a prescription dose of 30 Gy in three fractions. Four intensity-modulated radiotherapy (IMRT) plans were systematically generated for each patient by varying the number of fields (9 vs. 15) and minimum MU/segment (15 vs. 5): 9FL-IMRT (9 fields, 15 MU/segment), 9FS-IMRT (9 fields, 5 MU/segment), 15FL-IMRT (15 fields, 15 MU/segment), and 15FS-IMRT (15 fields, 5 MU/segment). Plan quality, treatment efficiency, and delivery accuracy were assessed based on dose distributions optimized using structure-based bulk electron density assignment, with patient-specific quality assurance (QA) evaluated via global gamma passing rate (GPR) using 3%/2 mm, 2%/2 mm, and 2%/1 mm criteria. To assess robustness against density-related spatial uncertainties, each plan was reoptimized after simulating skull misalignment with random translational offsets within ±2 mm. These reoptimized plans were recalculated on original computed tomography (CT) datasets using voxel-based electron density as the reference standard. Robustness was quantified by comparing dose distributions between bulk density-based and voxel-based recalculations through GPR analysis using 3%/2 mm, 2%/2 mm, 2%/1 mm, and 1%/1 mm criteria.ResultsAll four IMRT configurations provided clinically acceptable plans with similar target coverage and no significant differences in conformity, gradient, or homogeneity indices (all P>0.05). Although some dose parameters for normal brain tissue reached statistical significance (overall PConclusionsThe 15FL-IMRT configuration provided an optimal balance among plan quality, deliverability, and robustness of dose calculation, supporting its adoption as the preferred planning approach for MR-guided HSRT in patients with solitary medium-sized BMs.

Related in the same topics