Comparison of the performance of two thermoluminescent dosimetry systems for the personal dose-equivalent Hp(10) measurement.

Lu F, Li W, Jin J, Li K, Zhou J.

Abstract

The performance of two thermoluminescent dosimetry systems (RGD-3D and RE2000) manufactured in China and Finland was compared. Both of these dosimetry systems demonstrated satisfactory results as their performance met the requirements of the standard. The two dosimetry systems showed similar performance in the energy response. The RGD-3D dosimetry system performed better in nonlinear response, minimum detectable level and blind sample tests, whereas the RE2000 dosimetry system showed better stability.