Treatment planning system and beam data validation for the ZAP-X: A novel self-shielded stereotactic radiosurgery system.
Srivastava SP, Jani SS, Pinnaduwage DS, Yan X, Rogers L, Barranco FD (+2 more)
Abstract
Purpose To evaluate the treatment planning system (TPS) performance of the ZAP-X stereotactic radiosurgery (SRS) system through nondosimetric, dosimetric, and end-to-end (E2E) tests. Methods A comprehensive set of TPS commissioning and validation tests was developed using published guidelines. Nondosimetric validation tests included information transfer, computed tomography-magnetic resonance (CT-MR) image registration, structure/contouring, geometry, dose tools, and CT density. Dosimetric validation included comparisons between TPS and water tank/Solid Water measurements for various geometries and beam arrangements and end-to-end (E2E) tests. Patient-specific quality assurance was performed with an ion chamber in the Lucy phantom and with Gafchromic EBT3 film in the CyberKnife head phantom. RadCalc was used for independent verification of monitor units. Additional E2E tests were performed using the RPC Gamma Knife thermoluminescent dosimeter (TLD) phantom, MD Anderson SRS head phantom, and PseudoPatient gel phantom for independent absolute dose verification. Results CT-MR image registrations with known translational and rotational offsets were within tolerance ( 95% (5%/2 mm) for all four targets. For our first 58 patients, film-based patient-specific quality assurance has resulted in an average %GP of 98.7% (range, 94-100%) at 2%/2 mm. Conclusions Core ZAP-X features were found to be functional. On the basis of our results, point-dose and planar measurements were in agreement with TPS calculations using multiple phantoms and setup geometries, validating the ZAP-X TPS beam model for clinical use.