Saxsons Group — India's trusted nuclear medicine, radiotherapy, oncosurgery, dosimetry and cyclotron supplier since 1986
Peer-ReviewedPubMedReviewCited 50×Radiation safetyThe British journal of radiology · 2011

Personal dosimetry for interventional operators: when and how should monitoring be done?

Martin CJ.

Abstract

ObjectiveAssessment of the potential doses to the hands and eyes for interventional radiologists and cardiologists can be difficult. A review of studies of doses to interventional operators reported in the literature has been undertaken.MethodsDistributions for staff dose to relevant parts of the body per unit dose-area product and for doses per procedure in cardiology have been analysed and mean, median and quartile values derived. The possibility of using these data to provide guidance for estimation of likely dose levels is considered.ResultsDose indicator values that could be used to predict orders of magnitude of doses to the eye, thyroid and hands from interventional operator workloads have been derived, based on the third quartile values, from the distributions of dose results analysed.ConclusionDose estimates made in this way could be employed in risk assessments when reviewing protection and monitoring requirements. Data on the protection provided by different shielding and technique factors have also been reviewed to provide information for risk assessments. Recommendations on the positions in which dosemeters are worn should also be included in risk assessments, as dose measurements from suboptimal dosemeter use can be misleading.

Relevant Saxsons product