Dosimetric evaluation of the effect of dental implants in head and neck radiotherapy.

Ozen J, Dirican B, Oysul K, Beyzadeoglu M, Ucok O, Beydemir B.

Abstract

ObjectiveThe aim of the study was to examine the dose enhancement from scattered radiation at bone-dental implant interfaces during simulated head and neck radiotherapy.Study designFour cylindrical titanium dental implants with 3 different sizes and lengths were implanted into a human mandible in 4 different positions. Ionization measurements for 6 MV X, 25 MV X, and Co-60 gamma rays were done. Thermoluminescent dosimeter (TLD 100 ) chips were used to measure radiation dose enhancement due to the scattered electrons from titanium and electronic disequilibrium at the tissue-metal interface.ResultsThe results showed that for Co-60, there is a 21% maximum increase in dose to alveolar mandibular bone at the close proximity to the titanium. For 6-MV x-rays the dose enhancement increase was almost the same or slightly lower than for Co-60, while for 25-MV high-energy x-rays, dose enhancement was lower than that of others. This increase in dose enhancement fell off rapidly and became insignificant at 2 mm from the interface.ConclusionTotal dose that may lead to osteoradionecrosis risk of the mandible is slightly but not significantly affected by the scattered dose of the dental implants of lower jaw in the radiation field exposed to 3 different radiation beams.