Medical Physicist's Comparison · myOSL™ Chip vs MOSFET
A side-by-side comparison for the medical physicist evaluating an in-vivo dosimetry system. Three panels cover detector lifetime, system capacity and operating cost. Sources are cited inline so every number can be audited against the manufacturer specification or the peer-reviewed literature.
myOSL™ Chip
Reusable, with ~2 % sensitivity drift then stable
Each BeO chip is erasable and re-usable. Characterised sensitivity drift is approximately −2 % over the first 0–15 Gy cumulative dose, after which sensitivity is essentially stable to 32 Gy (the upper bound of the cited study). Reuse cycles are limited only by handling damage to the housing, not by saturation of the detector.
Source: Kowalski J et al., JACMP 26(4):e70057 (2025) — characterisation, commissioning and clinical evaluation of a commercial BeO OSL system.
MOSFET dosimeter
Replaced every 200–400 Gy cumulative
MOSFETs accumulate charge in the gate oxide with every irradiation. Once the cumulative service dose reaches the manufacturer-specified threshold — typically 200–400 Gy depending on the model — the dosimeter must be replaced (the threshold-voltage shift no longer tracks dose linearly).
Source: mobileMOSFET system documentation, Best Medical / RPDinc; characterisation literature (Cygler & Scarantino, AAPM 2009).
myOSL™ Chip
No fixed channel limit
myOSL Chips are passive — each is read out on the bench, after irradiation, by the handheld myOSL reader. As many chips as you have are read sequentially. There is no hardware ceiling on the number of in-vivo points per fraction; only operator time on the reader after the fact.
Source: Manufacturer workflow specification (handheld myOSLchip reader).
MOSFET dosimeter
Up to 5 MOSFETs per reader module
The mobileMOSFET Wireless Dosimetry System reader accepts up to 5 MOSFETs or one Linear 5ive Array per module. Scaling beyond 5 requires additional reader modules and transceivers — five concurrent dose points is the per-module ceiling.
Source: mobileMOSFET Wireless Dosimetry System, 5 Dose Points, 1 Reader Module — RPDinc product specification.
myOSL™ Chip
Reader + reusable chips — capital + minimal consumable
After the reader is purchased, the per-measurement cost is essentially zero until physical damage takes a chip out of service. Linearity, fading and angular response are characterised on every reuse via the type-test protocol.
Source: Manufacturer service model.
MOSFET dosimeter
Reader + recurring detector replacements
Each MOSFET has a finite cumulative-dose service life (200–400 Gy). A clinic running routine in-vivo dosimetry across multiple fractions and patients reaches that ceiling on a predictable cadence — detector replacement becomes a recurring consumable line on the QA budget.
Source: mobileMOSFET service documentation.
| Metric | myOSL™ Chip | MOSFET dosimeter |
|---|---|---|
| Detector lifetime | Reusable; −2 % drift 0–15 Gy, stable to 32 Gy | 200–400 Gy then replace |
| Simultaneous in-vivo points | No fixed limit (sequential read) | Up to 5 per reader module |
| Operating cost | Capital + minimal consumable | Capital + periodic detector replacement |
| Linearity (clinical range) | r² = 0.9999 to 10 Sv | Saturation depends on oxide thickness; ~0.1–5 Gy or 5+ Gy per device |
| Peer-reviewed validation | Kowalski 2025 + Davis 2025 (JACMP, multi-institution) | AAPM TG-191 OSL primary; MOSFET literature established since 1990s |
Every row is anchored to a panel above. The peer-reviewed papers cited for myOSL Chip live on the Knowledge Hub.
Sources cited on this page
myOSL™ Chip
Sibling posts in the myOSL Chip family.
For Medical Physicist
Linearity to 10 Sv, fading at 3 months, energy & angular response — the manufacturer type-test data in one page.
Read this Medical PhysicistTBI needs 8–10+ in-vivo points across a long fraction. Where does the myOSL Chip line up with the workflow — and what does the peer-reviewed evidence say?
Read this