Physicist's Notes · ART Phantom
End-to-end treatment-chain QA runs the same workflow as a patient treatment, with dosimeters in place of body tissue. Seven steps from CT-simulation through delivery to dossier filing. This post walks each step, the dosimeter-to-task mapping for the five measurement modalities, and how the dossier fits AAPM TG-119 (commissioning) and TG-218 (patient-specific) expectations.
Seven-step cycle
Image the phantom
CT-simulate the loaded phantom with the same protocol used for patient imaging — slice thickness, kV, mA, reconstruction algorithm matched.
Use the standard patient-imaging protocol. The CT dataset feeds the treatment planning system as if it were a patient.
Plan the treatment
Contour the phantom-internal target volumes (placed via TLD / dosimeter pin positions) and OARs. Plan a clinically realistic IMRT / VMAT treatment.
The plan should match the clinical complexity the QA is verifying — head-and-neck, prostate, breast, lung, etc.
Insert dosimeters
Load the chosen dosimeters (TLD, OSL, MOSFET, ion chamber, film) into the slice positions that correspond to the planned target and OAR points.
TLD positions stored in the standard phantom map; clinical-physics group has the position-to-anatomy lookup table.
Deliver the plan
Set the phantom up on the linac couch as if it were the patient — couch indexing, lasers aligned to phantom-marked iso-centre. Deliver the full treatment plan.
Couch indexing matters. Misaligned couch position is a common source of measured-dose discrepancy.
Read the dosimeters
Remove the dosimeters and read them on the corresponding reader (TLD reader for TLD, OSL reader for OSL, etc.). Record measured dose at each position.
Calibration of the reader against a NIST-traceable source is part of the dosimeter calibration cycle.
Compare against TPS
Plot measured dose vs TPS-predicted dose at each position. Accept within ± 3 % (gamma index 3 %/3 mm, 95 % pass rate for IMRT QA per TG-218).
Failure mode is not just absolute dose — relative geometry of error tells you whether the fault is in imaging, planning or delivery.
Document the result
File the result in the per-machine, per-modality QA dossier. Annual programmes file the result against the AAPM TG-119 / TG-218 commissioning baseline.
Documentation is what survives accreditation audit. The number alone is not enough; the comparison narrative is.
Source: AAPM TG-119 IMRT Commissioning; AAPM TG-218 Patient-Specific IMRT QA.
Dosimeter-to-task mapping
| Dosimeter | Use case | Phantom insertion |
|---|---|---|
| TLD chips | Point dose at target + OAR positions; low-cost per measurement | Pin-hole pluck-and-place; standard 3 cm × 3 cm or 1.5 cm × 1.5 cm grids |
| OSL dosimeters | Repeated-readout dosimetry across multiple QA cycles; lower fade error than TLD over weeks | Same pin holes as TLD; OSL holders standard accessory |
| MOSFET | High-resolution point dose for SRS / SBRT QA where mm-scale gradients matter | Small-diameter (1.5 mm) MOSFET fits the 5 mm pin hole |
| Ion chamber | Reference dosimetry — absolute dose at specific calibration points; gold-standard comparison | Larger pin-hole positions; specific chamber-compatible inserts required |
| Film | Continuous 2-D dose distribution between slices; gamma-map analysis | External assembly mode — open slice gaps to seat film sheets |
AAPM TG-119 — Commissioning
AAPM TG-218 — Patient-specific IMRT QA