Per-site build sheet · Thermoplastic Mask Family
Six workflows, one mask family —and the daily-imaging cadence each one runs on.
The mask family covers routine H&N IMRT, intracranial SRS, MR-Linac, paediatric brain SRT, trunk IGRT and 3-point re-irradiation. This page is the per-site build sheet — SKU + accessory kit + daily-imaging cadence + intrafractional-drift budget + CTV-to-PTV margin — anchored against the five peer-reviewed papers on the sibling knowledge hub.
Routine H&N IMRT / VMAT
SKU / configuration
RT-001 · S-Frame H&N mask
Accessory kit
Shoulder-retraction strap
Daily imaging cadence
Daily MVCT / CBCT verification
CTV-to-PTV margin
CTV-to-PTV 3-5 mm typical for standard H&N IMRT
Setup-accuracy anchor: Mean 6D translational setup 0.32 / −1.09 / 2.24 mm; rotational < 0.4° on every axis (You 2024 · MVCT · n = 920 images)
Chain: Mask moulds to patient at 65 °C simulation, hardens as it cools, indexes onto the S-frame headrest baseplate. Daily image guidance corrects the residual setup error to within CBCT verification tolerance.
Intracranial SRS / SRT
SKU / configuration
RT-002 · Full-face cranial + SRS mask
Accessory kit
SRS-frame + SRS bite-block
Daily imaging cadence
ExacTrac / CBCT pre-fraction + intra-fraction
CTV-to-PTV margin
CTV-to-PTV 1-2 mm at target under AAPM TG-42 < 2 mm overall targeting
Setup-accuracy anchor: Intrafractional 3D drift plateau: 0.21 ± 0.26 mm at 0-2 min → 0.53 ± 0.31 mm at 10 min under mask fixation (Mangesius 2019 · 96 fractions · 453 ExacTrac measurements)
Chain: Full-face shell + SRS bite-block hard-locks head geometry. Beam-on delivery ceiling per fraction is set by the intrafractional-drift plateau at ~10 min. If the plan exceeds that, split into two mask-on sessions or upgrade to a 3-point cushion + mask + bite-block configuration.
MR-Linac + MR-guided radiotherapy
SKU / configuration
RT-001 or RT-002 + MR-compatible fastener set
Accessory kit
MR-compatible baseplate + coil-compatible headrest
Daily imaging cadence
MR cine + daily-adaptive re-planning
CTV-to-PTV margin
Site-specific per MR-Linac protocol; typically tighter PTV enabled by daily adaptive planning
Setup-accuracy anchor: Systematic translational setup error < 1 mm across the multi-centre European MR-Linac cohort (Cuccia 2021 · Radiation Oncology · systematic review)
Chain: Polymer sheet is MR-safe by default; the non-ferromagnetic fastener set closes the loop for MR-Linac + MR-sim. Same indexed pitch on the MR-sim and MR-Linac; the CT-derived isocentre transfers without per-room recalibration.
Paediatric brain-tumour SRT
SKU / configuration
RT-002 · Paediatric full-face mask
Accessory kit
Paediatric SRS bite-block
Daily imaging cadence
CBCT + intra-fraction verification
CTV-to-PTV margin
CTV-to-PTV 2-3 mm typical for paediatric SRT (protocol-dependent)
Setup-accuracy anchor: Sized for paediatric anatomy; same low-temperature-activation moulding as the adult SRS configuration
Chain: The 65 °C activation temperature is well below the paediatric skin burn threshold, which is what makes the perforated-polymer moulding step usable on children. Alternative rigid stereotactic frames are not paediatric-appropriate under general anaesthesia; the mask is.
Trunk / thorax IGRT + IMRT
SKU / configuration
RT-003 · Full-body thermoplastic shell
Accessory kit
Indexed thorax baseplate + arm supports
Daily imaging cadence
Daily CBCT + kV / MV alignment
CTV-to-PTV margin
CTV-to-PTV 5-7 mm typical for lung + upper thorax IGRT (motion-management-dependent)
Setup-accuracy anchor: Best-in-class daily-CBCT alignment within 2 mm across 158 CBCT / 19 patients on the custom-headrest + mouth-bite mask configuration: 99.4 % AP / 100 % SI / 98.7 % RL (Mail 2024 · J Biol Methods)
Chain: Body-shell restrains trunk motion across the fraction series; pairs with the thermoplastic head mask (RT-001 or RT-002) for the H&N + nodal treatment plans that need dual immobilisation. Indexed onto the thorax baseplate from the Patient Positioning Solutions family.
3-point re-irradiation immobilisation
SKU / configuration
RT-001 or RT-002 + custom cushion + SRS bite-block
Accessory kit
Custom head cushion (mould-room-shaped)
Daily imaging cadence
Robotic-couch + CBCT correction
CTV-to-PTV margin
CTV-to-PTV 2 mm PTV margin enabled by the 3-point configuration
Setup-accuracy anchor: MD Anderson H&N SRT re-irradiation cohort: initial 3D vector 2.7 ± 1.4 mm corrected to 0.0 ± 0.3 mm axis-per-axis residual after robotic-couch adjustment (Wang 2016 · JACMP · 21 patients / 105 sessions)
Chain: The three-point (custom cushion + thermoplastic mask + SRS bite-block) configuration is the concrete answer for H&N re-irradiation where the standard configuration cannot deliver the < 2 mm PTV margin the CTV would otherwise leave for organs at risk.
Where next
Two more pages, depending on what you need.
Knowledge Hub →
Five peer-reviewed anchor citations
You 2024 · Mail 2024 · Olch 2014 (TG-176) · Mangesius 2019 (SRS drift) · Cuccia 2021 (MR-Linac). Every DOI open-access.
Product page →
Family specs + variant matrix
RT-001 / 002 / 003 SKUs, frame compatibility, MR-compatible fastener set, Accessories Matrix.